SAN-EI AM Solar Simulator —
New Standard for Measurement and Evaluation

From April 1, 2025, SAN-EI will evaluate all Solar Simulators based on “IEC 60904-9_2020” and “SAN-EI 2025-04” standards.
Those Standards also applies to the maintenance required to maintain the performance of SAN-EI Solar Simulators.

  • Based on requests from customers in various fields, we created this in-house standard in April 2025 for the purpose of evaluating the accuracy of Solar Simulators “under conditions not specified in conventional standards.”
  • “SAN-EI 2025-04” was created with reference to “IEC 60904-9_2020”.
  • The content will be updated according to changes in international standards and customer requirements. (the name will also be updated accordingly.)
  • “AM1.5G 300-1200nm” is based on “IEC 60904-9_2020”.
  • AM1.5G 300-1800nm and 300-2200nm” complies with “SAN-EI 2025-04”.
  • 300-1200nm, 300-1800nm and 300-2200nmat AM0” comply with “SAN-EI 2025-04“.
  • “Measurement and Evaluation” of AM1.5G and AM0 is performed as follows.

* The parts in red in this document are specified in “SAN-EI 2025-04“. The rest complies with “IEC 60904-9_2020”.

AM1.5G

Setting “Reference Irradiance” for Measurement
  • Measurement Receiver : Pyranometer or Reference Cell
    When “Effective Irradiation Surface” is □40mm to □50mm, we use Pyranometer.
    When “Effective Irradiation Surface” is □70mm to □400mm, we use Reference Cell. 
  • Measurement Position (Working Distance)
    Measurement Receiver is placed so that its light receiving surface (sensor surface) is at Working Distance.
  • Measurement Position (on Irradiation Surface)
    Center of Effective Irradiation Surface
measurement position

①  Wait until Lamp turns on and becomes “Stable” (20 to 30 minutes).
②  Using Measuring Receiver, measure Irradiance at 5 points shown in the figure.
③  Measured Irradiance is calculated using the following formula:




Maximum−Minimum
Maximum+Minimum


× 100 (%)

④  Adjust Lamp Position so that the calculated value is 1.5% or less.
⑤  Adjust Lamp Current so that Center Measurement Value becomes Constant of Measurement Receiver. 

measurement irradiation 5 points
  • Reference Irradiance
    1000W/m2(±5%)
    Irradiance is adjusted based on Sensitivity Constant of Measurement Receiver. 
  • Measurement Receiver : Pyranometer or Reference Cell
    When “Effective Irradiation Surface” is □40mm to 50mm, we use Pyranometer.
    When “Effective Irradiation Surface” is □70mm to 400mm, we use Reference Cell.

  • Measurement Position (Working Distance)
    Measurement Receiver is placed so that its light receiving surface (sensor surface) is at Working Distance.

  • Measurement Position (on Irradiation Surface)
    Center of Effective Irradiation Surface

  • Reference Irradiance
    1000W/m2(±5%)
    Irradiance is adjusted based on Sensitivity Constant of Measurement Receiver.

measurement position
  • Short Term Instability (STI)
    Measured 30 minutes after Lamp turned on
    Measurement Interval: 1/1000 seconds
    Measurement Time (number of times) : 1 second (1000 times)
    Measurement and Recording using Measurement Receiver and Data Logger
  • Long Term Instability (LTI)
    Measured 30 minutes after Lamp turned on
    Measurement Interval: 1/2 seconds
    Measurement Time (number of times) : 1 hour (7200 times)
    Measurement and Recording using Measurement Receiver and Data Logger
  • Extract Minimum and Maximum Output Voltage ​​from Measurement Receiver

  • Calculate “Temporal Instability” using the following formula.




Maximum−Minimum
Maximum+Minimum


× 100 (%)

  • Create a chart

  • “Temporal Instability” calculated using the above formula is evaluated based on the table below.

Short Term Instability (STI) Evaluation Grade
≤ 0.25 % A+
≤ 0.5 % A
≤ 2 % B
≤ 10 % C
Long Term Instability (LTI)Evaluation Grade
≤ 1 %A+
≤ 2 %A
≤ 5 %B
≤ 10 %C
  • Measurement Receiver : Pyranometer or Reference Cell
    When “Effective Irradiation Surface” is □40mm to 50mm, we use Pyranometer.
    When “Effective Irradiation Surface” is □70mm to 400mm, we use Reference Cell.

  • Measurement Position (Working Distance)
    Measurement Receiver is placed so that its light receiving surface (sensor surface) is at Working Distance.

  • Measurement Position (on Irradiation Surface)
    Center of Effective Irradiation Surface

  • Reference Irradiance
    1000W/m2(±5%)
    Irradiance is adjusted based on Sensitivity Constant of Measurement Receiver.

measurement position
  • Measurements are carried out based on the table below.
  • Measurement Receiver is placed so that its light receiving surface is at Working Distance.
  • Measurement Position (on Irradiation Surface)

    Measure 25 to 400 points as shown in the table below.

    (Note) Align the edge of the light receiving surface of Measurement Receiver with the edge of Effective Irradiation Surface. 

Effective Irradiation Surface SizeMeasurement ReceiverPhotosensitive Area SizeNumber of Measurement Points
□ 40 × 40 mmPyranometer□ 8 mm5 × 5         25
□ 50 × 50 mm□ 10 mm
□ 70 × 70 mmReference
Battery
□ 14 mm8 × 8         64
□ 80 × 80 mm□ 16 mm
□ 100 × 100 mm□ 20 mm10 × 10         100
□ 160 × 160 mm11 × 11         121
□ 200 × 200 mm13 × 13         169
□ 220 × 220 mm15 × 15         225
□ 250 × 250 mm20 × 20         400
□ 300 × 300 mm 
□ 400 × 400 mm 
measurement position
irradiation uniformity surface
  • The reading on Measurement Receiver(output voltage) will fluctuate slightly, so record Intermediate Value.

  • Extract Minimum and Maximum Output Voltage ​​from Measurement Receiver

  • Calculate “Irradiance Uniformity” using the following




Maximum−Minimum
Maximum+Minimum


× 100 (%)

  • Create a chart formula

  • “Irradiance Uniformity” calculated using the above formula is evaluated based on the table on the right.

Irradiance UniformityEvaluation Grade
≤ 1 %A+
≤ 2 %A
≤ 5 %B
≤ 10 %C
  • Measurement Receiver : Pyranometer or Reference Cell
    When “Effective Irradiation Surface” is □40mm to 50mm, we use Pyranometer.
    When “Effective Irradiation Surface” is □70mm to 400mm, we use Reference Cell.

  • Measurement Position (Working Distance)
    Measurement Receiver is placed so that its light receiving surface (sensor surface) is at Working Distance.

  • Measurement Position (on Irradiation Surface)
    Center of Effective Irradiation Surface

  • Reference Irradiance
    1000W/m2(±5%)
    Irradiance is adjusted based on Constant of Measurement Receiver.

measurement position
  • Using Spectrometer, measure and record Irradiance in the range 300-2200 nm to 1 nm increments.

  • Measurement Position is
    – When Effective Irradiation Area is □40mm to □100mm, only Center in the figure below.
    – When Effective Irradiation Area is □160mm to □400mm, 4 points in the figure below.

  • Except for Center, the light receiving surface edge of Spectroscope is made to contact Effective Irradiation Surface Edge.

effective irradiation area
  • Organize the data for each of 4 measurement points. (When Effective Irradiation Area is □40mm-□100mm, only Center.)

  • Calculate Interval Sum for each Wavelength.

  • Calculate Sum of each Wavelength Range as shown in the table below.

Wavelength Range to be evaluated (AM1.5G)
300–1200 nm300–1800 nm300–2200 nm
Wavelength RangeWavelength RangeWavelength Range
300–470 nm300–458 nm300–443 nm
470–561 nm458–536 nm443–506 nm
561-657 nm536–616 nm506–570 nm
657-772 nm616–703 nm570–637 nm
772-919 nm703–809 nm637–708 nm
919-1200 nm809–970 nm708–797 nm
 970–1208 nm797–900 nm
 1208–1800 nm900–1061 nm
  1061–1302 nm
  1302–2200 nm
  • Calculate Percentage for each Wavelength Range using the following formula.




Sum of Wavelength Range
Sum of All Wavelength Range


× 100 (%)

(NOTE) Calculate all Wavelength Ranges.

  • The following formula is used to calculate “Spectrum Match”

Value of the above formula (%) Values ​​in the following table (%) × 100 (%)

(NOTE) Calculate for each Wavelength Range.

Wavelength Range to be evaluated (AM1.5G)
300–1200 nm300–1800 nm300–2200 nm
Wavelength RangeRatioWavelength RangeRatioWavelength RangeRatio
300–470 nm16.61 %300–458 nm12.52 %300–443 nm9.98 %
470–561 nm16.74 %458–536 nm12.54 %443–506 nm9.93 %
561–657 nm16.67 %536–616 nm12.45 %506–570 nm9.97 %
657–772 nm16.63 %616–703 nm12.45 %570–637 nm9.99 %
772–919 nm16.66 %703–809 nm12.51 %637–708 nm9.96 %
919–1200 nm16.69 %809–970 nm12.49 %708–797 nm10.04 %
Total 100.00 %970–1208 nm12.47 %797–900 nm10.00 %
 1208–1800 nm12.56 %900–1061 nm10.03 %
 Total 100.00 %1061–1302 nm10.03 %
  1302–2200 nm10.08 %
  Total 100.00 %
  • Extract Minimum and Maximum values ​​from “Spectral Matches” for each Wavelength Range.

  • Extract Minimum and Maximum values ​​at each Measurement Position (1 or 4).

  • Minimum and Maximum of all measurements is taken as “Spectral Match” for that product.

  • Range (Minimum to Maximum) of “Spectral Match” determined above is evaluated based on the table below.

Wavelength Range to be evaluated (AM1.5G)
300–1200 nm 300–1800 nm 300–2200 nm
Spectral Match Evaluation Grade Spectral Match Evaluation Grade Spectral Match Evaluation Grade
92.5107.5 % A++
87.5112.5 % A+ 87.5112.5 % S+ 87.5112.5 % SS+
75125   % A 75125   % S 75125   % SS
60140   % B
40200   % C

AM0

Setting “Reference Irradiance” for Measurement
  • Measurement Receiver : Pyranometer or Reference Cell
    When “Effective Irradiation Surface” is □40mm to □50mm, we use Pyranometer.
    When “Effective Irradiation Surface” is □70mm to □400mm, we use Reference Cell. 
  • Measurement Position (Working Distance)
    Measurement Receiver is placed so that its light receiving surface (sensor surface) is at Working Distance.
  • Measurement Position (on Irradiation Surface)
    Center of Effective Irradiation Surface
measurement position

①  Wait until Lamp turns on and becomes “Stable” (20 to 30 minutes).
②  Using Measuring Receiver, measure Irradiance at 5 points shown in the figure.
③  Measured Irradiance is calculated using the following formula:




Maximum−Minimum
Maximum+Minimum


× 100 (%)

④  Adjust Lamp Position so that the calculated value is 1.5% or less.
⑤  Adjust Lamp Current so that Output Voltage of Measuring Receiver at Center is 1.2 times Sensitivity Constant. 

measurement irradiation 5 points
  • Reference Irradiance

    1348W/m2(±5%) —-This value is defined as “1 SUN of AM O”. ※❶
    Adjust Irradiance so that Output Voltage of Measurement Receiver is 1.2 times Sensitivity Constant. ※❷

※❶ Reason: why “1 SUN of AM O” is “1348W/m2”

  • It is well known that “1 SUN of AM1.5G” is “1000W/m2”.
  • This is “Total Radiant Energy Value from 280 to 4000 nm” based on “ASTM G173-03.”
    (”ASTM G173-03″ is data on AM1.5G Radiant Energy.)
  • Similarly, “Total Radiant Energy from 280 to 4000 nm” of “ASTM E-490” is “1348 W/m2.”
    (”ASTM E-490″ is data for AM 0 Radiant Energy.)

※❷ Reason for setting “1.2 times Sensitivity Constant of Measurement Receiver as 1 SUN of AM 0”

  • There is no dedicated Measuring Device for Measuring AM 0 Light.
  • Reference Cells and Pyranometers can only measure in 400-1100nm range.
  • Total Radiant Energy Value of “ASTM G173-03” in “400-1100 nm” range is “759 W/m2”.
  • Total Radiant Energy Value of “ASTM E-490” in “400-1100 nm” range is “908 W/m2”.
  • 「908W/m2」÷「759W/m2」=1.20
  • Radiant Energy of AM 0 at “400-1100 nm” is 1.2 times that of AM1.5G.
  • Therefore, “Radiant Energy that is 1.2 times Sensitivity Constant of Measuring Receiver” is defined as “1348 W/m2.”

    (NOTE) It only applies to Light with “Spectral Match of 75-125%.”
  • Measurement Receiver : Pyranometer or Reference Cell
    When “Effective Irradiation Surface” is □40mm to 50mm, we use Pyranometer.
    When “Effective Irradiation Surface” is □70mm to 400mm, we use Reference Cell.

  • Measurement Position (Working Distance)
    Measurement Receiver is placed so that its light receiving surface (sensor surface) is at Working Distance.

  • Measurement Position (on Irradiation Surface)
    Center of Effective Irradiation Surface

  • Reference Irradiance
    1348W/m2(±5%)
    Adjust Irradiance(Lamp Current) so that Output Voltage of 
    Measurement Receiver is 1.2 times Sensitivity Constant.

measurement position
  • Short Term Instability (STI)
    Measured 30 minutes after Lamp turned on
    Measurement Interval: 1/1000 seconds
    Measurement Time (number of times) : 1 second (1000 times)
    Measurement and Recording using Measurement Receiver and Data Logger
  • Long Term Instability (LTI)
    Measured 30 minutes after Lamp turned on
    Measurement Interval: 1/2 seconds
    Measurement Time (number of times) : 1 hour (7200 times)
    Measurement and Recording using Measurement Receiver and Data Logger
  • Extract Minimum and Maximum Output Voltage ​​from Measurement Receiver

  • Calculate “Temporal Instability” using the following formula.




Maximum−Minimum
Maximum+Minimum


× 100 (%)

  • Create a chart

  • “Temporal Instability” calculated using the above formula is evaluated based on the table below.

Short Term Instability (STI) Evaluation Grade
≤ 0.25 % A+
≤ 0.5 % A
≤ 2 % B
≤ 10 % C
Long Term Instability (LTI)Evaluation Grade
≤ 1 %A+
≤ 2 %A
≤ 5 %B
≤ 10 %C
  • Measurement Receiver : Pyranometer or Reference Cell
    When “Effective Irradiation Surface” is □40mm to 50mm, we use Pyranometer.
    When “Effective Irradiation Surface” is □70mm to 400mm, we use Reference Cell.

  • Measurement Position (Working Distance)
    Measurement Receiver is placed so that its light receiving surface (sensor surface) is at Working Distance.

  • Measurement Position (on Irradiation Surface)
    Center of Effective Irradiation Surface

  • Reference Irradiance
    1348W/m2(±5%)
    Adjust Irradiance(Lamp Current) so that Output Voltage of Measurement Receiver is 1.2 times Sensitivity Constant.

measurement position
  • Measurements are carried out based on the table below.
  • Measurement Receiver is placed so that its light receiving surface is at Working Distance.
  • Measurement Position (on Irradiation Surface)

    Measure 25 to 400 points as shown in the table below.

    (Note) Align the edge of the light receiving surface of Measurement Receiver with the edge of Effective Irradiation Surface. 

Effective Irradiation Surface SizeMeasurement ReceiverPhotosensitive Area SizeNumber of Measurement Points
□ 40 × 40 mmPyranometer□ 8 mm5 × 5         25
□ 50 × 50 mm□ 10 mm
□ 70 × 70 mmReference
Cell
□ 14 mm8 × 8         64
□ 80 × 80 mm□ 16 mm
□ 100 × 100 mm□ 20 mm10 × 10         100
□ 160 × 160 mm11 × 11         121
□ 200 × 200 mm13 × 13         169
□ 220 × 220 mm15 × 15         225
□ 250 × 250 mm20 × 20         400
□ 300 × 300 mm 
□ 400 × 400 mm 
measurement position
irradiation uniformity surface
  • The reading on Measurement Receiver(output voltage) will fluctuate slightly, so record Intermediate Value.

  • Extract Minimum and Maximum Output Voltage ​​from Measurement Receiver

  • Calculate “Irradiance Uniformity” using the following




Maximum−Minimum
Maximum+Minimum


× 100 (%)

  • Create a chart formula

  • “Irradiance Uniformity” calculated using the above formula is evaluated based on the table on the right.

Irradiance UniformityEvaluation Grade
≤ 1 %A+
≤ 2 %A
≤ 5 %B
≤ 10 %C
  • Measurement Receiver : Pyranometer or Reference Cell
    When “Effective Irradiation Surface” is □40mm to 50mm, we use Pyranometer.
    When “Effective Irradiation Surface” is □70mm to 400mm, we use Reference Cell.

  • Measurement Position (Working Distance)
    Measurement Receiver is placed so that its light receiving surface (sensor surface) is at Working Distance.

  • Measurement Position (on Irradiation Surface)
    Center of Effective Irradiation Surface

  • Reference Irradiance
    1348W/m2(±5%)
    Adjust Irradiance(Lamp Current) so that Output Voltage of Measurement Receiver is 1.2 times Sensitivity Constant.

measurement position
  • Using Spectrometer, measure and record Irradiance in the range 300-2200 nm to 1 nm increments.

  • Measurement Position is
    – When Effective Irradiation Area is □40mm to □100mm, only Center in the figure below.
    – When Effective Irradiation Area is □160mm to □400mm, 4 points in the figure below.

  • Except for Center, the light receiving surface edge of Spectroscope is made to contact Effective Irradiation Surface Edge.

effective irradiation area
  • Organize the data for each of 4 measurement points. (When Effective Irradiation Area is □40mm-□100mm, only Center.)

  • Calculate Interval Sum for each Wavelength.

  • Calculate Sum of each Wavelength Range as shown in the table below.

Wavelength Range to be evaluated (AM0)
300–1200 nm300–1800 nm300–2200 nm
Wavelength RangeWavelength RangeWavelength Range
300–448 nm300–437 nm300–421 nm
448–538 nm437–516 nm421–489 nm
538-636 nm516–600 nm489–557 nm
636-756 nm600–696 nm557–629 nm
756-927 nm696–820 nm629–714 nm
927-1200 nm820–991 nm714–820 nm
 991–1257 nm820–957 nm
 1257–1800 nm957–1151 nm
  1151–1461 nm
  1461–2200 nm
  • Calculate Percentage for each Wavelength Range using the following formula.




Sum of Wavelength Range
Sum of All Wavelength Range


× 100 (%)

(NOTE) Calculate all Wavelength Ranges.

  • The following formula is used to calculate “Spectrum Match”

Value of the above formula (%) Values ​​in the following table (%) × 100 (%)

(NOTE) Calculate for each Wavelength Range.

Wavelength Range to be evaluated (AM0)
300–1200 nm300–1800 nm300–2200 nm
Wavelength RangeRatioWavelength RangeRatioWavelength RangeRatio
300-448 nm16.67 %300–437 nm12.53 %300–421 nm10.02 %
448-538 nm16.56 %437–516 nm12.46 %421–489 nm10.02 %
538-636 nm16.65 %516–600 nm12.46 %489–557 nm9.96 %
636-756 nm16.27 %600–796 nm12.42 %557–629 nm9.94 %
756-927 nm16.84 %796–820 nm12.45 %629–714 nm10.05 %
927-1200 nm16.99 %820–991 nm12.52 %714–820 nm10.04 %
Total 100.00 %991–1257 nm12.51 %820–957 nm9.99 %
 1257–1800 nm12.64 %957–1151 nm10.02 %
 Total 100.00 %1151–1461 nm10.04 %
  1461–2200 nm9.93 %
  Total 100.00 %
  • Extract Minimum and Maximum values ​​from “Spectral Matches” for each Wavelength Range.

  • Extract Minimum and Maximum values ​​at each Measurement Position (1 or 4).

  • Minimum and Maximum of all measurements is taken as “Spectral Match” for that product.

  • Range (Minimum to Maximum) of “Spectral Match” determined above is evaluated based on the table below.

Wavelength Range to be evaluated (AM0)
300–1200 nm300–1800 nm300–2200 nm
Spectral MatchEvaluation GradeSpectral MatchEvaluation GradeSpectral MatchEvaluation Grade
92.5107.5 %A++  
87.5112.5 %A+87.5112.5 %S+87.5112.5 %SS+
75125   %A75125   %S75125   %SS
60140   %B 
40200   %C