SAN-EI AM Solar Simulator

AM Solar simulator

BSW-70D1-100SUN

By replacing the AM filter, the unit supports both AM1.5G and AM0.(Specifications are shown below.)



Optical PerformanceIEC 60904-9Actual Value
300–1200nm  Spectral Match *87.5–112.5%A+90– 110%
Irradiance Uniformity≤ 30%——1.0 – 1.5%
Temporal Instability of IrradianceSTI≤ 0.25%A+0.14 – 0.23%
LTI≤ 1%A+0.2 – 0.6%

Specification
AM-dependent
Air Mass FilterAM1.5GAM 0
Standard Irradiance (=1SUN)
280–4000nm
1000W/m²1348W/m²
1 SUN Available Time
(This is not guaranteed value.)
≥ hours≥ hours
Common Specifications
Initial Maximum Irradiance≥ W/m²
Irradiance adjustment rangeabout 40–100%
Effective Irradiated Areaφ 70mm
Working Distance310mm
Irradiation directionDownward
Collimation Half Angleabout 2 degrees
Declination Angleabout 4 degrees
Lamp Model3000W Xenon Lamp
External Irradiation Operation
(shutter: Open/Close)
Contact ON/OFF (Open/Close)
Input voltage200–240VAC
50/60Hz, 5000W (Max.)
Packaging Information
(This is a reference value.)
92cm×80cm×192cm, 180kg

82cm×70cm× 63cm, 59kg