SAN-EI AM Solar Simulator

AM Solar simulator

XHS-220S1

By replacing the AM filter, the unit supports both AM1.5G and AM0.(Specifications are shown below.)



Optical PerformanceIEC 60904-9Actual Value
300–1200nm  Spectral Match *92.5–102.5%A++96– 104%
Irradiance Uniformity≤ 2%A1.0 – 1.5%
Temporal Instability of IrradianceSTI≤ 0.25%A+0.14 – 0.24%
LTI≤ 1%A+0.2 – 0.6%

Specification
AM-dependent
Air Mass Filter AM1.5G AM 0
Standard Irradiance (=1SUN) 280–4000nm 1000W/m² 1348W/m²
1 SUN Available Time (This is not guaranteed value.) ≥1000 hours
Common Specifications
Initial Maximum Irradiance ≥1400W/m²
Irradiance adjustment range about 50–100%
Effective Irradiated Area □ 220mm × 220mm
Working Distance 470mm
Irradiation direction Downward
Collimation Half Angle about 2 degrees
Declination Angle about 2.5 degrees
Lamp Model 1600W Xenon Lamp 750W Halogen Lamp
External Irradiation Operation (shutter: Open/Close) Contact ON/OFF (Open/Close)
Input voltage 200–240VAC 50/60Hz, 3600W (Max.)
Packaging Information (This is a reference value.) 100cm×123cm×133cm, 160kg
73cm× 85cm×105cm, 98kg
48cm× 63cm× 56cm, 13kg

Product Quote

Items you select