SAN-EI AM Solar Simulator

AM Solar simulator

XES-300S1

By replacing the AM filter, the unit supports both AM1.5G and AM0.(Specifications are shown below.)



xes 300s1

Optical PerformanceIEC 60904-9Actual Value
300–1200nm  Spectral Match *87.5–112.5%A+91– 109%
Irradiance Uniformity≤ 2%A1.0 – 1.3%
Temporal Instability of IrradianceSTI≤ 2%B0.5 – 0.8%
LTI≤ 2%A+0.6 – 1.2%

Specification
AM-dependent
Air Mass Filter AM1.5G AM 0
Standard Irradiance (=1SUN) 280–4000nm 1000W/m² 1348W/m²
1 SUN Available Time (This is not guaranteed value.) ≥1500 hours ≥1000 hours
Common Specifications
Initial Maximum Irradiance ≥1400W/m²
Irradiance adjustment range about 50–100%
Effective Irradiated Area □ 300mm × 300mm
Working Distance 924mm
Irradiation direction Downward
Collimation Half Angle 1.2-1.8 degrees
Declination Angle 4.0-5.0 degrees
Lamp Model 3000W Xenon Lamp
External Irradiation Operation (shutter: Open/Close) Contact ON/OFF (Open/Close)
Input voltage 200–240VAC 50/60Hz, 4500W (Max.)
Packaging Information (This is a reference value.) 116cm×74cm×116cm, 146kg
82cm×68cm× 62cm, 47kg