Solar simulator □100mm
XES-100S1 (A0)
The SAN-EI ELECTRIC XES-100S1 portable solar simulator delivers A+ rated performance (IEC 60904-9) with precise 1000W/m² output across a 100mm² area. Its 300W xenon lamp lasts 2500+ hours and supports multi-directional irradiation and optional filters. Featuring exclusive Arc Monitor alignment, this compact solution enables accurate sunlight simulation for PV testing, research, and quality control.
FEATURE
- Top-class approximation to sunlight (A+/A+/A/A+)
- 1 SUN lasted 2500 hours over (recorded value)
- Lightweight and compact, easy to carry.
- Original “Arc Monitor” makes optical adjustments very easy
- Easy to replace or add filters (AM0/Dimming/other) *Option
- Changeable direction of irradiation (Up/Down/Left/Right)
SPECIFICATION
| Product Model | XES-100S1 |
| Air Mass Filter | AM0 |
| Standard Irradiance | 1348W/m² (1 SUN) *280-4000nm |
| Effective Irradiated Area | □ 100mm × 100mm |
| Working Distance | 400mm |
| Collimation Half Angle | About 3 degrees |
| Declination Angle | About 5 degrees |
What's New?
| Item | Past | Now |
|---|---|---|
| Temporal Instability of Irradiance | Class A | Class A+ |
| Spectral Match (IEC 60904-9_2020) | Class A | Class A+ |
| Wavelength range to be evaluated | 400-1100nm | 300-1200nm 300-1800nm 300-2200nm |
Optical Performance
| Standard | IEC 60904-9_2020 | SAN-EI Performance | ||
|---|---|---|---|---|
| Reference Value | Class | Actual Value | ||
| Temporal Instability of Irradiance | STI | ≤ 0.5% | A | 0.2 – 0.3% |
| LTI | ≤ 1% | A+ | 0.5 – 0.8% | |
| Irradiance Uniformity | ≤ 2% | A | 1.0 – 1.5% | |
| Spectral Match (300-1200nm) | 87.5 – 112.5% | A+ | 92 – 108% | |
Light Source
| Lamp Model | 300W Xenon Lamp |
| Initial Maximum Irradiance | ≥ 1500W/m² |
| 1 SUN Available Time | ≥ Approximately 2500 hours |
Optical Control
| Irradiation ON/OFF A | By opening and closing the shutter |
| How to open and close the shutter B | Main unit switch / external switch |
| Irradiance Adjustment C | By adjusting the lamp current (100–50%) |
- The shutter can be opened and closed remotely using the remote connector (B).
- The shutter opens when contact 7 and contact 8 connect.
- The shutter closes when the connection between contact 7 and contact 8 opens.
Operational Safety
| Automatic forced air cooling after lights off | Fan runs for 20 minutes, then stops |
| Inter Lock (If the cover is not attached properly) | Cannot light on or forcibly light off |
| Overheating prevention mechanism | Forcibly light off by thermostat |
| Lamp Hour Meter D | Total lighting time (resettable) |
Electrical Specifications
| Rated Input | 100 – 120VAC, 50/60Hz, 1φ 480W (Max.) |
| 200 – 240VAC, 50/60Hz, 1φ 480W (Max.) | |
| Voltage Endurance | Input-Output / F.G: 1000VAC-1min. |
| Output / F.G: 500VAC-1min. | |
| Insulation Resistance | Input-Output / F.G: DC 500V above 20MΩ |
| Output / F.G: DC 500V above 20MΩ |
Environmental Conditions
| Temperature | 20–30°C (RT) * The temperature change should be within 5°C |
| Humidity | 20–85% (RH) * No condensation |
Dimensions and Weight (Light source)
Dimensions and Weight (Controller)
- The warranty applies to products that have been delivered for less than one year and have been used for less than 500 hours ( lamp lighting time).
(This does not apply to lamps.)
- The following cannot be guaranteed even during the warranty period.
- Accidents, breakdowns, or malfunctions caused by use or operation not in accordance with the instruction manual
- Accidents, breakdowns, or defects that occur after modifications are made without our consent.
- Accidents, breakdowns, and defects that occur when using lamps without our shipping inspection records.
In addition, as a general rule, we will not accept investigations or repairs for accidents, breakdowns, or defects that occur